BÜCHI Labortechnik AG
Editing a method | 8
Operation Manual FatExtractor E-500
59/88
The following parameters are set automatically:
Rinse heating level (without analyte protection only)
Drying time
Drying heating level
Navigation path
➔
u
Navigate to the
[Method]
menu via the navigation path.
u
Tap the name of the method that you wish to edit.
ð
The display shows the view
Method
.
u
Tap the function
[Edit]
on the function bar.
u
Tap the action
[SmartDrying]
.
ð
The function is enabled.
8.12
Setting the drying time of a method
NOTICE
Risk of beaker damage or deterioration of the analyte due to incorrect
settings of the drying step.
Incorrect drying step settings can damage the analyte and the beaker.
u
Changing the default settings is at your own risk.
The duration of the drying time of a method depends on the solvent volume.
Navigation path
➔
u
Navigate to the
[Method]
menu via the navigation path.
u
Tap the name of the method that you wish to edit.
ð
The display shows the view
Method
.
u
Tap the function
[Edit]
on the function bar.
u
Select the action
[Drying time]
with the navigation control.
u
Turn the navigation control until the required value.
u
Tap the function
[OK]
on the function bar.
u
Tap the function
[Save]
on the function bar.
ð
The time is saved.
8.13
Setting the drying heating level of a method
The default setting for the drying heating level is depending on the selected solvent.
Change the drying heating level to compensate fluctuations in ambient conditions.