6 | Operating the interface
BÜCHI Labortechnik AG
46/88
Operation Manual FatExtractor E-500
Action
Option
Explanation
[Actors]
Change settings
The following settings can be
changed:
Valves:
Soxhlet valves See Chapter 11.4
"Draining solvent from Extraction
glass chamber (Soxhlet only)",
page 76
Tank valves See Chapter 11.3
"Draining solvent from the instru-
ment", page 75
Cooling water valve On/Off
Heater On/Off
Pump On/Off
Lift current
u
Tap up or down.
ð
Display shows the actual
current consumption.
Beeper On/Off
Chiller On/Off (option)
Submenu System information
The submenu
[System information]
contains details of the connected devices and in-
formation on network connection diagnosis.
Action
Option
Explanation
[Extraction unit]
View
The following information on the Ex-
traction Unit is available:
Serial number
Firmware version
Operating hours
Board temperature
Code
[Interface]
View
The following information on the in-
terface is available:
Serial number
Firmware version
Operating hours
Board temperature
Code
30 V Operating voltage
5 V Operating voltage
Submenu History QR code
Shows the history list of the data export QR codes to export data. SeeChapter 9.6
"Exporting extraction parameters", page 70