5 | Installation
BÜCHI Labortechnik AG
28/88
Operation Manual FatExtractor E-500
NOTICE
Risk of property damage and diminished performance due to use of un-
suitable power cables.
The power supply cables supplied with the product by BUCHI precisely match the re-
quirements of the device. If other power cables that do not meet those requirements
are used, the device may be damaged and/or its performance diminished.
u
Use only the power supply cables supplied with the product or ordered separately
from BUCHI.
u
If using any other power supply cables, make sure that they match the specifica-
tions on the type plate.
u
Make sure that all connected devices are grounded.
u
Make sure that the power plug is freely accessible at all times.
u
Insert the power cable into the connection labeled
Power IN
on the back of the in-
strument.
u
Insert the power plug into the power socket.
5.4
Securing against earthquakes
The instrument has an earthquake fixing point to protect the device against falling.
1
Fig. 11: Securing against earthquakes
1
Lashing mount
u
Tie the lashing mount to a fixed point using strong cord or a wire.
5.5
Connecting the coolant hose
There are two possibilities for coolant supply:
Cooling water valve Order no. 031356
BUCHI Recirculating Chiller F-305 or F-308