8 | Editing a method
BÜCHI Labortechnik AG
58/88
Operation Manual FatExtractor E-500
WARNING
Setting the heating level too high
The consequences are device damage, serious injuries or death.
u
Use the pre-set heating levels.
u
Adjust the heating level to compensate the ambient conditions.
Navigation path
➔
u
Navigate to the
[Method]
menu via the navigation path.
u
Tap the name of the method that you wish to edit.
ð
The display shows the view
Method
.
u
Tap the function
[Edit]
on the function bar.
u
Select the action
[Rinse heating level]
with the navigation control.
u
Turn the navigation control until the required value.
u
Tap the function
[OK]
on the function bar.
u
Tap the function
[Save]
on the function bar.
ð
The heating level is saved.
8.10
Setting the number of drains for a method (Hot Extraction only)
During each drain 3 - 4 mL of solvent is drained into the solvent tank.
The level in the beaker is lowered to finally rinse the sample without touching the ex-
tract.
Navigation path
➔
u
Navigate to the
[Method]
menu via the navigation path.
u
Tap the name of the method that you wish to edit.
ð
The display shows the view
Method
.
u
Tap the function
[Edit]
on the function bar.
u
Select the action
[No. of drains]
with the navigation control.
ð
The display shows a dialog with the selectable values.
u
Select the value you wish to use with the navigation control.
u
Tap the function
[OK]
on the function bar.
u
Tap the function
[Save]
on the function bar.
ð
The value is saved.
ð
The dialog box closes.
8.11
Setting SmartDrying (Soxhlet and Economic Continuous
Extraction only)
Smart-Drying automatically sets the parameters for the drying step.