5 | Installation
BÜCHI Labortechnik AG
30/88
Operation Manual FatExtractor E-500
5.7
Connecting the coolant tubes to the condenser tank bottle
Precondition:
R
Make sure that the instrument is not con-
nected to the power supply.
u
Install the inlet tube onto the connection
marked
CW TANK OUT
.
u
Attach the inlet tube in place with a hose clip.
u
Install the drain tube onto the connection
marked
CW TANK IN
.
u
Attach the drain tube in place with a hose clip.
u
Attach the inlet tube onto the connection
marked
CW IN
.
u
Attach the drain tube onto the connection
marked
CW OUT
.
u
Make sure that the solvent collector circuit is
complete. See Chapter 14.1 "Schematics",
page 81
5.8
Preparing the tank bottle
NOTICE
Risk of broken parts by using tools
Using tools for preparing the instrument can break parts.
u
Tighten all parts with your hands.
u
Switch the
On/Off
master switch to Off.
u
Put the bottle cap on the bottle.