BÜCHI Labortechnik AG
Editing a method | 8
Operation Manual FatExtractor E-500
57/88
Open time
Level sensor posi-
tion
Explanation
[mid]
middle
The sample structure is easily releasing
the solvent.
[long]
top
The sample structure is easily releasing
the solvent.
bottom
The sample structure is slowly releasing
the solvent.
[max]
top
The Soxhlet LSV chamber is used.
The sample structure is slowly releasing
the solvent.
Navigation path
➔
u
Navigate to the
[Method]
menu via the navigation path.
u
Tap the name of the method that you wish to edit.
ð
The display shows the view
Method
.
u
Tap the function
[Edit]
on the function bar.
u
Select the action
[Sox valve open time]
with the navigation control.
ð
The display shows a dialog with the selectable values.
u
Tap the function
[OK]
on the function bar.
u
Tap the function
[Save]
on the function bar.
ð
The Soxhlet valve open time is saved.
8.8
Setting the rinse time of a method (Soxhlet and Hot Extraction
only)
Enter the duration of the extraction step in minutes.
at least 0 minutes
maximum 5940 minutes / 99 hours
Navigation path
➔
u
Navigate to the
[Method]
menu via the navigation path.
u
Tap the name of the method that you wish to edit.
ð
The display shows the view
Method
.
u
Tap the function
[Edit]
on the function bar.
u
Select the action
[Rinse time]
with the navigation control.
u
Turn the navigation control until the required value.
u
Tap the function
[OK]
on the function bar.
u
Tap the function
[Save]
on the function bar.
ð
The time is saved.
8.9
Setting the rinse heating level of a method (Soxhlet and Hot
Extraction only)
The default setting for the rinse heating level is depending on the selected solvent.
Change the rinse heating level to compensate ambient conditions.