43
Z31914_00_01
Assembling the Tray with Rotors
i
Notice: The tray must be inserted with the barcode on the left side.
i
Notice: The slot must be on the left side, so that the barcode labels can be read.
i
Notice: When working in the mode
48 Rotors no cooling device
, use a non-cooled
rotor container. The container must be set in the last position to the right, as shown in the
following figure. Do not forget to use the adapter plate for room temperature to fix the
rotor container. The adapter plate must be inserted before inserting the tray. The plate
must be inserted so that the text "FRONT" is in the front.
Figure 9.1
Adapter Plate for Room Temperature in Last Position to the Right
9.
Cover the cooling rack with
the cap.
Содержание SamplePro hr-MAS
Страница 1: ...Version Innovation with Integrity NMR Spectroscopy SamplePro hr MAS 001 Sample Changer User Manual ...
Страница 6: ...6 Z31914_00_01 Table of Contents ...
Страница 38: ...38 Z31914_00_01 Installation and Initial Commissioning ...
Страница 44: ...44 Z31914_00_01 Assembling the Tray with Rotors ...
Страница 57: ...57 Z31914_00_01 Operation Figure 10 14Activity History ...
Страница 58: ...58 Z31914_00_01 Operation ...
Страница 62: ...62 Z31914_00_01 Preventative Maintenance ...
Страница 65: ...65 Z31914_00_01 13 EC Declaration of Conformity ...
Страница 66: ...66 Z31914_00_01 EC Declaration of Conformity ...
Страница 68: ...68 Z31914_00_01 Contact ...
Страница 70: ...70 Z31914_00_01 Warning Signs ...
Страница 72: ...72 Z31914_00_01 Figures ...
Страница 74: ...74 Z31914_00_01 Tables ...
Страница 76: ...76 Z31914_00_01 Glossary ...
Страница 80: ...80 Z31914_00_01 Index ...
Страница 81: ......