Environment and Site Survey Measurement
48
H157655_1_008
7.1.5
Floor Vibration Guidelines: Bruker Nano-C and Nano-D
ADI damping system (Air Spring Damped Isolators with vertical damping)
Frequency
Maximum Velocity in
Vertical Direction
Maximum Velocity in
Horizontal Direction
1-5 Hz
3 µm/s
1.6 µm/s
5-20 Hz
3…12 µm/s
1.6…6.4 µms
>20 Hz
12 µm/s
6.4 µm/s
Table 7.3:
ADI Damping System (Elastomeric Isolators)
Figure 7.5: Example of the Bruker Tolerance Level for ADI Dampers. Resolution <0.125 Hz, 1…200 Hz,
> 30 min, Average Measurement 1 (displayed in velocity)
Содержание AVANCE
Страница 6: ...Contents vi H157655_1_008 ...
Страница 24: ...System Components 24 H157655_1_008 ...
Страница 32: ...Magnet Access and Rigging 32 H157655_1_008 ...
Страница 38: ...Ceiling Height Requirements 38 H157655_1_008 ...
Страница 42: ...Magnetic Stray Fields 42 H157655_1_008 6 3 Stray Field Plots Figure 6 1 Example of a Stray Field Plot ...
Страница 72: ...Utility Requirements 72 H157655_1_008 ...
Страница 82: ...Floor Plan 82 H157655_1_008 ...
Страница 96: ...CryoProbe and Other Accessories 96 H157655_1_008 ...
Страница 102: ...Contact 102 H157655_1_008 ...
Страница 104: ...List of Figures 104 H157655_1_008 ...
Страница 109: ...H157655_1_008 109 ...
Страница 110: ...Bruker Corporation info bruker com www bruker com Order No H157655 ...