Safety
18
H157655_1_008
2.8.1
Fire Department Notification
It is recommended that the magnet operator introduce the fire department and/or local
authorities to the magnet site. It is important that these organizations be informed of the
potential risks of the magnet system, e.g. that much of the magnetic rescue equipment
(oxygen-cylinders, fire extinguishers, axe's etc.) can be hazardous close to the magnet
system. In addition, their expertise and experience can be invaluable in creating an
Emergency plan.
• In a NMR laboratory use only non-magnetic fire extinguishers.
• Breathing equipment which uses oxygen tanks made out of magnetic material can be life
threatening when used close to a magnet system which is energized.
• During a quench helium gas escaping from the system must not be mistaken for smoke.
Instruct the fire department and technical service not to „extinguish“ the magnet system
with water. The outlet valves could freeze over the quench valves eventually do not close
again.
• Laboratory windows which are accessible during an emergency must be clearly marked
with warning signs, visible from the outside.
Содержание AVANCE
Страница 6: ...Contents vi H157655_1_008 ...
Страница 24: ...System Components 24 H157655_1_008 ...
Страница 32: ...Magnet Access and Rigging 32 H157655_1_008 ...
Страница 38: ...Ceiling Height Requirements 38 H157655_1_008 ...
Страница 42: ...Magnetic Stray Fields 42 H157655_1_008 6 3 Stray Field Plots Figure 6 1 Example of a Stray Field Plot ...
Страница 72: ...Utility Requirements 72 H157655_1_008 ...
Страница 82: ...Floor Plan 82 H157655_1_008 ...
Страница 96: ...CryoProbe and Other Accessories 96 H157655_1_008 ...
Страница 102: ...Contact 102 H157655_1_008 ...
Страница 104: ...List of Figures 104 H157655_1_008 ...
Страница 109: ...H157655_1_008 109 ...
Страница 110: ...Bruker Corporation info bruker com www bruker com Order No H157655 ...