Floor Plan
H157655_1_008
73
9
Floor Plan
9.1
Magnet Location
When locating the magnet, certain considerations must be made with regards to the
laboratory environment:
• To increase magnet homogeneity, the magnet should be located away from permanent
iron structures such as support beams in walls and floors. Reference:
• To increase temperature stability, the magnet should not be placed in direct sunlight or
near any artificial heat source. The magnet should also not be placed under or in close
proximity to air-vents or in an area that experiences air drafts. Air should not be blown
directly down or towards the NMR magnet.
• When possible, avoid a situation where a significant stray field (>5 G / 0.5 mT) extends
into adjacent rooms.
• There should be free access to the magnet from all sides.
Figure 9.1: Top View of an Ascend Magnet
Содержание AVANCE
Страница 6: ...Contents vi H157655_1_008 ...
Страница 24: ...System Components 24 H157655_1_008 ...
Страница 32: ...Magnet Access and Rigging 32 H157655_1_008 ...
Страница 38: ...Ceiling Height Requirements 38 H157655_1_008 ...
Страница 42: ...Magnetic Stray Fields 42 H157655_1_008 6 3 Stray Field Plots Figure 6 1 Example of a Stray Field Plot ...
Страница 72: ...Utility Requirements 72 H157655_1_008 ...
Страница 82: ...Floor Plan 82 H157655_1_008 ...
Страница 96: ...CryoProbe and Other Accessories 96 H157655_1_008 ...
Страница 102: ...Contact 102 H157655_1_008 ...
Страница 104: ...List of Figures 104 H157655_1_008 ...
Страница 109: ...H157655_1_008 109 ...
Страница 110: ...Bruker Corporation info bruker com www bruker com Order No H157655 ...