![Bruker AVANCE Скачать руководство пользователя страница 27](http://html.mh-extra.com/html/bruker/avance/avance_user-manual_2820950027.webp)
Magnet Access and Rigging
H157655_1_008
27
Crate Size (m)
Magnet Transport Dimensions (m)
(for transport to the magnet room)
System
(MHz/mm)
L
D
H
Width
Uncrated
Height
Uncrated w/
o Pallet
Jack*
Height
Uncrated
with Pallet
Jack**
400/89 Ascend Aeon
1.15
1.35
2.08
0.85
1.53
1.64
500/89 Ascend Aeon
1.15
1.36
2.02
0.95
1.74
1.85
600/89 Ascend Aeon
1.15
1.35
2.08
0.85
1.53
1.64
700/89 Ascend Aeon
1.50
1.30
2.14
0.85
1.53
1.64
* Measured from magnet bottom plate to helium tower - this is the absolute minimum
height!
** The heights listed with pallet jack assume that the floor is level, thus the magnet needs
only to be jacked up approx. 2 cm for transport. If the floor is uneven, the magnet may
need to be jacked up accordingly, which could add as much as 9 cm to the transport
height.
Table 4.2:
Door Dimensions: Wide Bore 89 mm
4.3.2
Magnet Stand Transport Dimensions
Door Dimensions for Magnet Stand & Accessories Access
Accessories Crate Size - including stand if
applicable (m)
Magnet Stand
L
D
H
400/54 Ascend Aeon F85 570 ADI
0.97
0.76
1.20
400/54 Ascend Aeon F85 700 ADI
0.97
0.76
1.20
400/54 Ascend Aeon F85 570 API
0.97
0.76
1.20
400/54 Ascend Aeon F85 700 API
0.97
0.76
1.20
500/54 Ascend Aeon F85 570 ADI
0.97
0.76
1.20
500/54 Ascend Aeon F85 700 ADI
0.97
0.76
1.20
500/54 Ascend Aeon F85 570 API
0.97
0.76
1.20
500/54 Ascend Aeon F85 700 API
0.97
0.76
1.20
600/54 Ascend Aeon F95 700 API
1.77
0.97
1.11
700/54 Ascend Aeon F95 700 API
1.77
0.97
1.11
Pallet is integrated in crate. Add 2-10 cm for pallet jack depending on floor quality. Allow
at least 1 cm clearance on the sides and above the crate.
Table 4.3:
Door Dimensions for Magnet Stand: Standard Bore 54 mm
Содержание AVANCE
Страница 6: ...Contents vi H157655_1_008 ...
Страница 24: ...System Components 24 H157655_1_008 ...
Страница 32: ...Magnet Access and Rigging 32 H157655_1_008 ...
Страница 38: ...Ceiling Height Requirements 38 H157655_1_008 ...
Страница 42: ...Magnetic Stray Fields 42 H157655_1_008 6 3 Stray Field Plots Figure 6 1 Example of a Stray Field Plot ...
Страница 72: ...Utility Requirements 72 H157655_1_008 ...
Страница 82: ...Floor Plan 82 H157655_1_008 ...
Страница 96: ...CryoProbe and Other Accessories 96 H157655_1_008 ...
Страница 102: ...Contact 102 H157655_1_008 ...
Страница 104: ...List of Figures 104 H157655_1_008 ...
Страница 109: ...H157655_1_008 109 ...
Страница 110: ...Bruker Corporation info bruker com www bruker com Order No H157655 ...