![Bruker AVANCE Скачать руководство пользователя страница 61](http://html.mh-extra.com/html/bruker/avance/avance_user-manual_2820950061.webp)
Utility Requirements
H157655_1_008
61
8.3.4
Compressed Air System
When designing a suitable compressed air system the following points must be taken into
consideration:
• To prevent magnetic impurities from entering the magnet use only copper or stainless
steel lines. Do not use iron or steel pipes. Plastic piping is unsuitable where very low dew
points are required. Water vapor in the air will permeate plastic piping limiting minimum
dew points to typically -25°C.
• To avoid surges in the air pressure (e.g. during sample lift) install a container of 10-20
liters in the air supply line to act as a buffer. Locate the buffer after the dryers in the supply
line.
Buffer containers
must meet the appropriate safety requirements. They must have a
working pressure of 16 bar and be proofed up to 30 bar. Use tanks which are internally
coated with water and acid resistant material. This will prevent corrosion from impurities
such as SO
2
.
The three major components in a suitable compressed air supply line include the
compressor, dryer and appropriate filters:
Figure 8.1: Example of a Typical Dryer/Filter System Setup
Filter 1:
General purpose liquid and dust removal filter (0.1 mg/m
3
- 0.1 ppm, 1
micron).
Filter 2:
High-efficiency dust, liquid and aeresol filter (0.1 mg/m
3
- 0.01 ppm, 1
micron).
When using a dryer/filter system setup, the following questions should be addressed:
• Pressure loss in piping?
• Efficiency loss in dryer?
• Pressure loss in filter?
• What is the required pressure?
• What is the required flow rate?
Содержание AVANCE
Страница 6: ...Contents vi H157655_1_008 ...
Страница 24: ...System Components 24 H157655_1_008 ...
Страница 32: ...Magnet Access and Rigging 32 H157655_1_008 ...
Страница 38: ...Ceiling Height Requirements 38 H157655_1_008 ...
Страница 42: ...Magnetic Stray Fields 42 H157655_1_008 6 3 Stray Field Plots Figure 6 1 Example of a Stray Field Plot ...
Страница 72: ...Utility Requirements 72 H157655_1_008 ...
Страница 82: ...Floor Plan 82 H157655_1_008 ...
Страница 96: ...CryoProbe and Other Accessories 96 H157655_1_008 ...
Страница 102: ...Contact 102 H157655_1_008 ...
Страница 104: ...List of Figures 104 H157655_1_008 ...
Страница 109: ...H157655_1_008 109 ...
Страница 110: ...Bruker Corporation info bruker com www bruker com Order No H157655 ...