System Components
H157655_1_008
19
3
System Components
This section describes the types and functions of the various sub-systems that are delivered
as part of our AVANCE UltraStabilized NMR systems. These include the following:
•
Superconducting Magnet Components [
•
Console and Other System Components [
•
•
.
3.1
Superconducting Magnet Components
The superconducting magnet is a complex system producing a very strong, homogeneous,
and stable magnetic field as required for NMR. This section describes the various sub-
systems of the magnet system.
Magnet
:
The magnet system’s main component is a superconducting coil housed
in a cryostat. The cryostat consists of an outer vacuum enclosure, some
radiation shields and a liquid helium vessel.
The magnet uses liquid helium as cryogenic liquid. The magnet coil is
immersed in a sub-cooled liquid helium (~2 K) bath. An additional liquid
helium bath operating at a standard temperature of 4.2 K is located
above the sub-cooled helium section and is also housed in the outer
vacuum enclosure.
After the initial charging with electrical current, the magnet runs in
persistent mode. The current runs in a closed loop inside the system and
the magnet itself is no longer connected to a continuous power supply.
Pulse Tube
Cooler
:
The magnet system is equipped with one cryocooler (pulse tube type).
The cold head is mounted on top of the magnet. The rotary valve is
mounted on a column right next to the magnet.
The cryocooler re-liquefies helium that has been extracted by pumping
and virtually leads to cryogen consumption free operation.
Helium
Compressor
:
An oil-lubricated helium compressor is used to supply pressurized helium
gas for PTC operation. This compressor requires water cooling and
electrical power without any interruption.
Maintenance
: Magnet maintenance consists of refilling the system with cryogenic fluids
at defined time intervals.
Содержание AVANCE
Страница 6: ...Contents vi H157655_1_008 ...
Страница 24: ...System Components 24 H157655_1_008 ...
Страница 32: ...Magnet Access and Rigging 32 H157655_1_008 ...
Страница 38: ...Ceiling Height Requirements 38 H157655_1_008 ...
Страница 42: ...Magnetic Stray Fields 42 H157655_1_008 6 3 Stray Field Plots Figure 6 1 Example of a Stray Field Plot ...
Страница 72: ...Utility Requirements 72 H157655_1_008 ...
Страница 82: ...Floor Plan 82 H157655_1_008 ...
Страница 96: ...CryoProbe and Other Accessories 96 H157655_1_008 ...
Страница 102: ...Contact 102 H157655_1_008 ...
Страница 104: ...List of Figures 104 H157655_1_008 ...
Страница 109: ...H157655_1_008 109 ...
Страница 110: ...Bruker Corporation info bruker com www bruker com Order No H157655 ...