Floor Plan
76
H157655_1_008
Magnet
A
Maximum
Magnet
Diameter
(m)
B
Magnet Height
from the Floor
Including Stand
(m)
C
Overall
Footprint
Diameter
(m
2
)
Total Magnet
Weight incl.
Stand &
Cryogens
(kg)*
400/89 Ascend Aeon
0.90
2.86
2.25
~1050
500/89 Ascend Aeon
0.90
2.86
2.25
~1130
600/89 Ascend Aeon
0.90
2.86
2.25
~1370
700/89 Ascend Aeon
1.05
3.03
2.74
~1990
* With standard magnet stand, the weight with optional stands will vary.
Table 9.3:
Magnet Dimensions and Weight: Wide Bore 89 mm
9.4
Floor Types
Generally a
liquid nitrogen resistant floor material
must be used, such as PVC or wood
that has been painted or varnished. Unfinished wood must not be used as this will absorb
liquid nitrogen. This also implies that wood floors must be regularly maintained to help
prevent absorption.
Many of the system components contain highly sensitive electronic devices that must be
protected from
Electrostatic Discharge
(ESD) by proper floor covering and grounding
practices.
To prevent ESD damage in the magnet room, the system must be installed on an ESD
resistant flooring such as vinyl, and properly grounded. One of the most important
characteristics of an
ESD resistant floor
is its ability to conduct charges to ground. The
second most important aspect is its
anti-static property
.
9.5
Magnet Pits
When the magnet must be sited in an existing magnet pit, pay attention to the following:
• Special rigging equipment and a temporary platform to support and lower magnet inside
the pit.
• Continuous ventilation and emergency exhaust inside the pit (please refer to special notes
HVAC (Heating Ventilation Air Conditioning) [
and
Emergency Ventilation During Installation and Quenches [
).
• Magnet refills and access for transport dewars.
• Cable lengths.
• Siting the BCU-I or BCU-II cooling unit.
• Siting the CryoPlatform
TM
.
Consult your local Bruker Installation Engineer for details.
Содержание AVANCE
Страница 6: ...Contents vi H157655_1_008 ...
Страница 24: ...System Components 24 H157655_1_008 ...
Страница 32: ...Magnet Access and Rigging 32 H157655_1_008 ...
Страница 38: ...Ceiling Height Requirements 38 H157655_1_008 ...
Страница 42: ...Magnetic Stray Fields 42 H157655_1_008 6 3 Stray Field Plots Figure 6 1 Example of a Stray Field Plot ...
Страница 72: ...Utility Requirements 72 H157655_1_008 ...
Страница 82: ...Floor Plan 82 H157655_1_008 ...
Страница 96: ...CryoProbe and Other Accessories 96 H157655_1_008 ...
Страница 102: ...Contact 102 H157655_1_008 ...
Страница 104: ...List of Figures 104 H157655_1_008 ...
Страница 109: ...H157655_1_008 109 ...
Страница 110: ...Bruker Corporation info bruker com www bruker com Order No H157655 ...