![Bruker AVANCE Скачать руководство пользователя страница 37](http://html.mh-extra.com/html/bruker/avance/avance_user-manual_2820950037.webp)
Ceiling Height Requirements
H157655_1_008
37
Ceiling Height Requirements
The number in the following tables corresponds to a standard magnet configuration using a
minimum height stand. The ceiling height can increase up to 250 mm based on the model of
the magnet and stand. Refer to the table in Magnet Stand Transport Dimensions for
specification on the magnet model and stand.
Magnet Type
Special LHE Transfer Line
for Reduced Ceiling
Height (m)
Standard LHE Transfer
Line (m)
400/54 Ascend Aeon
2.66
3.04
500/54 Ascend Aeon
2.66
3.04
600/54 Ascend Aeon
3.00
3.39
700/54 Ascend Aeon
3.00
3.39
Table 5.2:
Minimum Ceiling Height Requirements: Standard Bore 54 mm
Magnet Type
Special LHE Transfer Line
for Reduced Ceiling
Height (m)
Standard LHE Transfer
Line (m)
400/89 Ascend Aeon
3.25
3.64
500/89 Ascend Aeon
3.25
3.64
600/89 Ascend Aeon
---
3.65
700/89 Ascend Aeon
3.40
3.79
Table 5.3:
Minimum Ceiling Height Requirements: Wide Bore 89 mm
Содержание AVANCE
Страница 6: ...Contents vi H157655_1_008 ...
Страница 24: ...System Components 24 H157655_1_008 ...
Страница 32: ...Magnet Access and Rigging 32 H157655_1_008 ...
Страница 38: ...Ceiling Height Requirements 38 H157655_1_008 ...
Страница 42: ...Magnetic Stray Fields 42 H157655_1_008 6 3 Stray Field Plots Figure 6 1 Example of a Stray Field Plot ...
Страница 72: ...Utility Requirements 72 H157655_1_008 ...
Страница 82: ...Floor Plan 82 H157655_1_008 ...
Страница 96: ...CryoProbe and Other Accessories 96 H157655_1_008 ...
Страница 102: ...Contact 102 H157655_1_008 ...
Страница 104: ...List of Figures 104 H157655_1_008 ...
Страница 109: ...H157655_1_008 109 ...
Страница 110: ...Bruker Corporation info bruker com www bruker com Order No H157655 ...