ENGLISH OUS indications
Page 32 of 50
OTH-SEM-IFU-OUS-0359 Rev I
Upon placement into the Charging Hub, the scanner switches into charging mode and initiates
wireless communications with the configured wireless network and then connects to the
Gateway Server. All communications are encrypted and secure.
Once the scanner has connected to the Gateway Server, all stored data sessions are uploaded
to the Gateway Server and then deleted from the Provizio SEM Scanner. There have been
occasional cases of the data not automatically uploading, this is usually because the Facility WIFI
has not initiated
–
a red X will appear
–
in which case:
•
Try simply removing and replacing the device into the Charging Hub
–
if the issue
persists
•
Ensure there are no known WIFI issues ongoing in Facility
–
if none
•
Refer the issue to your Facility IT support
–
they have access to a Guide to decode
the error and find a solution
If interrupted during the data transfer, the scanner will resend the interrupted and unsent data
the next time it is placed in the charger.
Turning the Scanner OFF
The scanner may be turned off by pressing and holding the action button.
After being turned off and then on, the scanner will re-start and begin a new patient scanning
session.