ENGLISH OUS indications
Page 45 of 50
OTH-SEM-IFU-OUS-0359 Rev I
Separation distance
The Provizio SEM Scanner is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The user of the Provizio SEM Scanner can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the Provizio SEM Scanner as recommended
below, according to the maximum output power of the communications equipment.
The recommended separation distance between portable and mobile RF communications
equipment and the Provizio SEM Scanner is listed in Table 15.
Table 15 Separation Distance
Separation distance according to frequency of transmitter (metres)
Rated maximum
output power of
transmitter (W)
150 kHz to 80 MHz
P
d
=
2
.
1
80 MHz to 800 MHz
P
d
=
2
.
1
800 MHz to 2.5 GHz
P
d
=
3
.
2
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter,
where P is the maximum output rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.