8 Reprocessing
42
BOWA-IFU-MN031-554-ERGO315R-S3-EN-20210219
8.7.
Packing
Sterilisation in the transportation packaging is not allowed.
The packaging must meet the following requirements:
•
EN (ANSI AAMI) ISO 11607 / EN 868-2...10 (formerly EN
868 / ANSI AAMI ISO 11607)
•
Suitable for steam sterilisation (resistant to temperatures
up to 137°C, sufficient steam permeability)
•
Regular maintenance (sterilisation container)
Check the instrument components for completeness.
Autoclave the handle only in the unlatched position.
Place the components in the reprocessing basket as shown in the
following figure:
Pack the reprocessing basket with the components in a suitable
disposable sterilisation package and/or a suitable sterilisation
container
.