8 Reprocessing
40
BOWA-IFU-MN031-554-ERGO315R-S3-EN-20210219
Suitable cleaning agents
Ensure that the selected cleaning agent system fulfils the following
requirements:
•
The cleaning agent is suitable for the vessel sealing
instrument and the chemicals used are compatible with
the instrument components
Follow the manufacturer’s instructions with regard to the concentration
and exposure time of the cleaning and disinfection products.
•
The manufacturer accepts no responsibility for the use of other
cleaning and disinfection products.
Cleaning and disinfection
1.
Place the reprocessing basket in the CDM and start the program.
2.
After the program is finished, remove the reprocessing basket
from the CDM.
3.
Detach the cleaning adapters in the reverse order of attachment
(see section 8.5.1, page 36).
4.
Use filtered compressed air to dry the instrument components
and remove any residual moisture.
NOTE
Risk of handle damage due to compressed air!
Restrict compressed air pressure to 3 bar or less for
handle drying.