6 Operation
BOWA-IFU-MN031-554-ERGO315R-S3-EN-20210219
27
Separating tissue
The tissue has been grasped by the jaws and is coagulated.
WARNING
Strong bleeding may occur if the grasped tissue is cut
before coagulation or vessel sealing!
With larger vessels, make several seals to the left and
right of the cutting site as necessary.
Before cutting, ensure that the tissue has been reliably
sealed.
Cut only in the sealed area.
11. To cut, pull the trigger and then release it.
The tissue will be cut through the middle of the sealed area.
6.4.
Removal
WARNING
Risk of patient injury due to damaged or detached parts!
Check the vessel sealing instrument after each use. All
parts must be present.
1.
Close the electrodes.
2.
Withdraw the vessel sealing instrument from the trocar sleeve.
6.5.
After use
WARNING
Defective, worn and contaminated electrodes may lead to
functional failure of the vessel sealing instrument!
Clean the electrodes regularly with a moist cloth.
Replace the jaw if the electrodes are damaged.
Replace the blade after every operation.
Prepare the vessel sealing instrument after use (see section 8,
page 32).