8 Reprocessing
34
BOWA-IFU-MN031-554-ERGO315R-S3-EN-20210219
8.2.
Soaking
Immediately after use, remove dirt residues from the open
jaw with a damp non-woven cloth.
Soak the disassembled vessel sealing instrument in cold
water or aldehyde-free disinfectant for at least 5 minutes
immediately after use, at the latest after 2 hours, and
remove coarse dirt residues with a plastic brush.
To soak the jaw, BOWA recommends a small tube filled with
liquid, e.g. the Leonardo-Wet-Set® from Interlock
Medizintechnik GmbH, which you push over the branches of
the jaw immediately after use.
The disinfectant used for soaking is intended solely for
personal protection and does not replace subsequent
disinfection.