3 Safety instructions
10
BOWA-IFU-MN031-554-ERGO315R-S3-EN-20210219
Clean and sterilise the ERGO 315R handle and the associated
electrode before each use.
Use only cleaning, disinfection and sterilisation methods that have
been validated for the specific devices and products concerned.
Comply with the validated parameters for each cycle.
Observe the applicable legal regulations in your country and the
hygiene regulations of the hospital.
To clean the instrument correctly, it is essential to remove the
replaceable blade.
Do not carry out gas sterilisation and hot air sterilisation, they can
damage the product.
•
To avoid combustion (thermal overload), observe the following points:
Make sure that all electrical plug connections are correctly
connected.
Do not stress the products beyond their mechanical limits (e.g. by
excessive bending, buckling or crushing).
The lowest maximum voltage of the product must not be exceeded.
•
In order to avoid the risk of injuries and electric shocks for the patient
and the operating personnel, ensure that the power supply is switched
off before connecting or removing the HF connection cable and
accessories to or from the electrosurgical device.
3.2.
Use-related
•
Only the Ligation and ARC SEAL modes may be used with the
BOWA ARC generators.
•
The electrode should only be activated with suitable visual control.
•
Do not use the product on vessels with a diameter greater than 7.0 mm.
•
Do not use vessel sealing for tissue groups with unknown content.
•
Proceed cautiously and maintain distance when performing sealing in
the vicinity of sensitive structures, such as nerves or ureter
•
Do not use too much tissue for sealing. The jaw must not be too full.
•
The reliability of sealing must be assessed by the attending physician
according to the nature of the vascularised tissue and the vessel
pathology (arteriosclerosis, aneurysms, vascularisation, etc.).