8 Reprocessing
32
BOWA-IFU-MN031-554-ERGO315R-S3-EN-20210219
8.
Reprocessing
WARNING
Risk of patient injury from non-sterile vessel sealing
instruments!
The vessel sealing instrument is not sterile when
delivered. Clean and sterilise the instrument before use.
Clean and sterilise the vessel sealing instrument before
each subsequent use.
Use only cleaning, disinfection and sterilisation methods
that have been validated for the specific devices and
products concerned.
Comply with the validated parameters for each cycle.
Observe the applicable legal regulations in your country
and the hygiene regulations of the hospital.
Danger of microbial transmission!
To clean the instrument correctly, it is essential to
remove the replaceable blade
.
Vessel sealing instruments must always be cleaned, disinfected and
sterilised before use. Effective cleaning and disinfection are essential for
effective subsequent sterilisation of the vessel sealing instruments.
1.
Ensure that only adequately validated device- and product-
specific methods are used for cleaning, disinfection and
sterilisation and that the validated parameters are complied with
in each cycle.
2.
Observe the applicable national legal regulations and the
hygiene regulations of the hospital or clinic.
BOWA-electronic GmbH & Co. KG assumes no warranty for malfunction of
the vessel sealing instruments in connection with disinfectants or the
method used, including the effectiveness of the disinfectant.