3 Safety instructions
12
BOWA-IFU-MN031-554-ERGO315R-S3-EN-20210219
3.2.1.
Patients with pacemakers
Malfunction or destruction of the pacemaker can endanger the life of the
patient or result in irreversible injuries to the patient.
Carry out a postoperative pacemaker check.
3.3.
EMC information
Medical electrical devices are subject to special EMC precautions,
so please observe the following directions.
The BOWA accessory is only intended to be connected to BOWA specified
HF devices.
Using the accessory with medical devices from other manufacturers
can result in higher emission levels or reduced interference immunity.
Never perform outpatient operations on patients with pacemakers.
In cases of patients with pacemakers, consult the cardiologist before
carrying out HF surgery.
Set the demand pacemaker to a fixed frequency.
Ensure that the pacemaker does not come into contact with the
HF electrode.
Keep a fully operational defibrillator within reach
NOTE
Combining medical devices is only safe if
the desired combination is allowed by the instructions for
use, and
the purpose and the interface specification of the devices
used in the combination allow it.
Strict attention must be given to the instructions for use and
interface specifications of medical devices used in
combination.