8 Reprocessing
BOWA-IFU-MN031-554-ERGO315R-S3-EN-20210219
35
8.3.
Manual removal of contamination
CAUTION
Risk of infection due to water spray and vapours from the
ultrasonic bath or with manual precleaning!
Wear a face mask and protective clothing.
Adequate ventilation is recommended.
NOTE
Risk of material damage to the jaw by scouring products
and metal brushes!
Never use scouring products or metallic brushes to
clean the vessel sealing instrument.
Recommended procedure
1.
Clean
all parts inside and outside, using the included accessory
brushes or other suitable plastic brushes.
Clean the blade groove of the jaw with the cleaning tip on the
blade holder.
2.
Rinse
all individual parts from the inside and outside after
brushing with a suitable rinsing gun for at least 15 s (5 pulses of
3 s) and 1.8
– 2 bar water pressure.
8.4.
Pretreatment in an ultrasonic bath
1.
Place all parts of the vessel sealing instrument in the ultrasonic
bath for at least 10 minutes at 35°C, e.g. with 1% neodic
MediClean forte (Dr. Weigert, Hamburg)
1
. Position instrument
parts with large surface areas in the ultrasonic bath such that
there are no acoustic shadows.
).
2.
Rinse all individual parts from the inside and outside after the
ultrasonic bath with a suitable rinsing gun for at least 15 s
(5 pulses of 3 s) and 1.8
– 2 bar water pressure.
1
Validation: for 10 minutes, 35°C ±2°C, 35 kHz, 0.3% neodic MediClean forte