11 Technical specifications
62
Operating Manual ARC 250 / 303
900-301_IFU-V2.2_11316-S0-20131023-EN
Programs
Number of programs in the device
100
Standard programs, factory set
Individually programmable
Program number and information shown on the
display
Monitoring of the neutral electrode
EASY
:
E
lectrode
A
pplication
Sy
stem
Indicator for one-piece or split electrode on the
front panel
Transition resistance between the individual
surfaces (CQM) of split neutral elecrodes shown
on the display
Lead resistance shown on the display when using
a one-piece neutral electrode (CM continuity
monitor)
Maximum resistance between the sections of split
electrodes (CQM)
999 Ω
Maximum allowable resistance between the
sections of split electrodes
220 Ω
Warning signal in the case of risk under the
neutral electrode
Visual, acoustic
Warning indicator as a text on the display:
Safety features
ISSys
:
I
ntegrated
S
afety
Sy
stem
Continuous monitoring of the HF leakage current
with error messages
Monitoring of the dosage, error message on the
display
Continuous self-test
Continuous status indication on the display
Display of operating errors on the display
Display of system errors on the display
Documentation
Recording and storage of data in the device
Error statuses
Operating errors
Retrieval of this data via the display
Содержание ARC 250
Страница 1: ...OPERATING MANUAL ELECTROSURGICAL UNIT...
Страница 2: ......
Страница 3: ...900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 3...
Страница 4: ...4 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...
Страница 5: ...900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 5...
Страница 8: ...Legend 8 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...
Страница 64: ...11 Technical specifications 64 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...
Страница 65: ...11 Technical specifications 900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 65...