7 Detecting and correcting faults
900-301_IFU-V2.2_11316-S0-20131023-EN
Operating Manual ARC 250 / 303
53
Error no.
(INF)
Signals/
System
events
Cause
Corrective measures
47
No power set for the activated output
Set the power for the activated output.
49
Connection socket activated via finger
and foot switches
At the end of activation a different
activation signal is pending for the
same socket
Activate the output via the finger or
foot switch one after the other.
61
Tissue contact is already pending
upon activation of the AUTOSTART
mode
Remove the instrument from the
tissue and re-activate AUTOSTART.
90
Papillotome or snare is not applied
Apply the papillotome or the snare.
91
Instrument for GastroCut application
defective or short circuit
Replace the Gastro instrument.
Select a new point of application.
92
Excessive low-resistance activation
(e.g. in saline)
Ensure that the instrument has
sufficient contact with the tissue for
coagulation or cutting before
activating the device.
100-104
EASY neutral electrode error
Check the neutral electrode and the
neutral electrode cable, see
section
Neutral electrode
monitoring
, page 30.
146
Internal error
Restart the device. If the error recurs,
contact the service center.
147
Internal error
Restart the device. If the error recurs,
contact the service center.
Содержание ARC 250
Страница 1: ...OPERATING MANUAL ELECTROSURGICAL UNIT...
Страница 2: ......
Страница 3: ...900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 3...
Страница 4: ...4 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...
Страница 5: ...900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 5...
Страница 8: ...Legend 8 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...
Страница 64: ...11 Technical specifications 64 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...
Страница 65: ...11 Technical specifications 900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 65...