Setting
Program
Gastro LOOP
6
Values measured on ohmic resistances
Output power P [W] versus rated load
resistance R [Ω] with "Gastro LOOP",
"Blend 0–9" settting, COAG phase
Values measured on ohmic resistances
Output power P [W] versus load
resistance R [Ω] with "Gastro LOOP",
"Blend 0–9" setting, CUT phase
Blend
Coag share
in W
Cut share
in W
0
36
370
1
36
370
2
36
370
3
36
370
4
36
370
5
36
370
6
36
370
7
36
370
8
36
370
9
36
370
Output power P [W] versus Blend
setting for "Gastro LOOP". Rated load
resistance: Coag 75 Ω, Cut 500 Ω
Blend
HF voltage [Vp]
0
800
1
800
2
800
3
800
4
800
5
800
6
800
7
800
8
800
9
800
HF output voltage U [Vp] versus
"Gastro LOOP" "Blend 0–9" (no load)
The short-term peak power exceeds 400 W, but this is permissible
because the mean value over 1 second is below 400 W.
Содержание ARC 250
Страница 1: ...OPERATING MANUAL ELECTROSURGICAL UNIT...
Страница 2: ......
Страница 3: ...900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 3...
Страница 4: ...4 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...
Страница 5: ...900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 5...
Страница 8: ...Legend 8 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...
Страница 64: ...11 Technical specifications 64 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...
Страница 65: ...11 Technical specifications 900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 65...