5 Preparation
900-301_IFU-V2.2_11316-S0-20131023-EN
Operating Manual ARC 250 / 303
37
5.3.4.
Assigning a foot switch output
1. Press the extra button on the dual foot pedal swiitch (e.g. for
"Bipolar Output"
31
).
– or –
Key/indicator for "Foot Switch"
25
.
The "Bipolar Foot Switch" indicator
25
will light up and an
alternating acoustic signal will be emitted for acoustic confirmation
of the switchover process. The blue foot switch pedal is now
activated for bipolar coagulation.
2. Press the key/display for AUTOSTART
26
for longer than
5 seconds.
The configured HF parameters will be saved and the symbol ->
appears on the display or main display
15
.
Pos : 35 /679-BOWA/HF- Ger äte/ARC 300/350/5 Vorbereitung/F unkti onspr üfung - Titel @ 6\mod_1297768711604_6.doc @ 48568 @ 2
5.4.
Functional test
Pos : 36 /679-BOWA/HF- Ger äte/ARC 200/250/300e/5 Vor ber eitung/F unkti ons prüfung @ 6\mod_1297336252619_6.doc @ 48300 @ 3333
5.4.1.
Autotest function
The HF device automatically carries out a self-test after being switched
on and a cyclical test during operation. If errors occur, see
section Detecting and correcting faults, page 51.
5.4.2.
Functional test execution
Perform the following functional test before putting the device into
service:
The accessories must be designed for the indicated maximum voltage.
1. Connect the neutral electrode and attach it to the patient’s arm.
The EASY neutral electrode indicator
19
/
20
changes to green.
2. Remove the neutral electrode.
The indicator
18
changes to red, acoustic signals sound, and the
EASY neutral electrode monitoring indicator
19
/
20
goes dark.
3. Press the surfaces of the neutral electrode against each other.
The EASY neutral electrode monitoring indicator
19/20
changes
back to green.
The neutral electrode used for this test may not later be used for an
operation.
4. Connect a monopolar HF handpiece to a monopolar socket
connector
29/30
and use the hand and foot switches to individually
activate "Cut" and "Coag".
Содержание ARC 250
Страница 1: ...OPERATING MANUAL ELECTROSURGICAL UNIT...
Страница 2: ......
Страница 3: ...900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 3...
Страница 4: ...4 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...
Страница 5: ...900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 5...
Страница 8: ...Legend 8 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...
Страница 64: ...11 Technical specifications 64 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...
Страница 65: ...11 Technical specifications 900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 65...