5 Preparation
900-301_IFU-V2.2_11316-S0-20131023-EN
Operating Manual ARC 250 / 303
35
3. Set up the HF device with sufficient clearance to other electronic
devices, see section EMC, page 86.
4. Position the HF device with the front of the device facing the
patient/surgeon.
5. Do not place any other objects on or above the HF device.
6. Do not set up the HF device on other devices.
7. Connect the power cable.
Pos : 32 /679-BOWA/HF- Ger äte/ARC 200/250/300e/5 Vor ber eitung/HF- Ger ät einschalten @ 7\mod_1309496096443_6.doc @ 52418 @ 2
5.2.
Switching on the HF device
Do not use the HF device if the indicator elements are not working. For
error correction, see section Detecting and correcting faults, page 51.
1. Switch the HF device on using the on/off switch
27
.
The HF device carries out a self-test: all indicator elements light up
and blink alternately with the display.
All modes that can be set in the menu programs are displayed
alternately. For more information about the setting options, see
section Menu programs, page 45.
2. Check whether all the indicators (
2
,
7
,
9
,
12
,
13
,
14
,
15
,
22
,
25
,
26
)
and LEDs (
1
,
8
,
18
,
19
,
20
,
21
) on the front panel light up and flash
alternately with the display.
3. Press any key to end the self-test.
The HF device is ready for operation.
The parameters of the most recently selected program appear on
the display.
A self-test is carried out if the device is switched off for longer than
15 to 20 seconds after being used. The default parameters of the
most recently selected program subsequently appear.
Pos : 33 /679-BOWA/HF- Ger äte/ARC 200/250/300e/5 Vor ber eitung/Z ubehör ansc hließ en @ 6\mod_1297066132449_6.doc @ 47914 @ 233
5.3.
Connecting instruments
Before connecting instruments, ensure that the following conditions
are met:
Combinations of accessories not mentioned in the
operating manual may be used only if they are explicitly
designed for the intended use. Performance features as
well as safety requirements must always be taken into
account.
The insulation of the accessories (e.g. HF cables,
instruments) must be sufficient for the maximum output
peak voltage (see IEC 60601-2-2 and IEC 60601-2-18).
Do not use accessories with defective insulation.
Содержание ARC 250
Страница 1: ...OPERATING MANUAL ELECTROSURGICAL UNIT...
Страница 2: ......
Страница 3: ...900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 3...
Страница 4: ...4 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...
Страница 5: ...900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 5...
Страница 8: ...Legend 8 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...
Страница 64: ...11 Technical specifications 64 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...
Страница 65: ...11 Technical specifications 900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 65...