7 Detecting and correcting faults
900-301_IFU-V2.2_11316-S0-20131023-EN
Operating Manual ARC 250 / 303
55
7.2.
Error Display in EASY monitoring
When working with a split neutral electrode, the following errors may
occur:
EASY monitoring
20
Cause
Indicator
Corrective measures
Blinks yellow
Significant increase in
resistance
Heat may develop under the
neutral electrode depending
on indication
–
Application need not be
terminated.
Switches from green to
continuous red
When the monopolar current
is activated, a significant
problem occurs
An acoustic signal sounds.
A corresponding "INF - xxx"
message appears on the
"Monopolar Cut" 7-segment
display
2
and the "Monopolar
Coag" 7-segment display
9
.
Check the neutral
electrode and the
neutral electrode
cable, see
section
Neutral
electrode
monitoring
, page 30
Check the neutral
electrode cable for
proper contact and
external damage.
Loosening electrode
An acoustic signal sounds. A
corresponding "INF – xxx"
message appears.
The device disables the
monopolar outputs to sockets
29/30
and
31
.
Correct the position of
the neutral electrode.
In the case of
continuing error
messages, replace
the neutral electrode.
Pos : 63 /679-BOWA/HF- Ger äte/ARC 300/350/8 Aufber eitung/UE_Aufbereitung @ 4\mod_1261040691288_6.doc @ 35166 @ 1
Содержание ARC 250
Страница 1: ...OPERATING MANUAL ELECTROSURGICAL UNIT...
Страница 2: ......
Страница 3: ...900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 3...
Страница 4: ...4 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...
Страница 5: ...900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 5...
Страница 8: ...Legend 8 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...
Страница 64: ...11 Technical specifications 64 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...
Страница 65: ...11 Technical specifications 900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 65...