5 Preparation
38
Operating Manual ARC 250 / 303
900-301_IFU-V2.2_11316-S0-20131023-EN
5. Check the settings on the display.
6. Now change to the bipolar output and connect bipolar tweezers.
7. Press the key/display for AUTOSTART
26
again and use the foot
switch to activate the bipolar output. Check the settings and
indicators in the bipolar section.
8. Press the key/display for AUTOSTART
26
again and use the foot
switch to activate the bipolar output. Check the settings and
indicators in the bipolar section.
5.4.3.
Actions in case of problems
Proceed as follows in case of functional problems:
1. Immediately disconnect the patient from the HF device.
2. Inspect the HF device and perform a functional test.
3. Report incidents and near-accidents to the German Federal
Institute for Medications and Medical Products in accordance with
Section 3 of the German Ordinance on the Installation, Operation
and Use of Medical Products (MPBetreibV). Observe the provisions
of the in-house reporting system in this regard.
4. Contact the service center,see section Technical service, page 59.
5.4.4.
EASY neutral electrode electrode monitoring
(EASY monitoring)
NOTE
Risk of incorrect application of the neutral electrode!
Ensure compliance with the specifications for correct application of
the neutral electrode with regard to size, adhesive properties and
full-surface contact of the complete electrode.
The following functions can be tested for split and one-piece neutral
electrodes:
Actual status
Output
Measure
Cable for one-piece or split
neutral electrode not connected
Indicators for EASY monitoring of split neutral
electrode
20
or EASY monitoring of one-piece
neutral electrode
19
do not light up
Plug in the cable for the
one-piece or split neutral
electrode.
Only the cable for the one-
piece or split neutral electrode
is connected
Indicators for EASY monitoring of split neutral
electrode
20
or EASY monitoring of one-piece
neutral electrode
19
light up
Check the cable for a short
circuit.
Cable with split neutral
electrode plugged in but not
attached to patient.
Indicators for EASY monitoring of split neutral
electrode
20
or EASY monitoring of one-piece
neutral electrode
19
light up
Check whether the electrode
is correctly attached to the
cable connection.
The cable for the split or one-
piece neutral electrode is
plugged in and attached to the
patient.
The indicators for EASY monitoring of the split
neutral electrode
20
or EASY monitoring of
the one-piece neutral electrode
19
light up
green, but the type of neutral electrode is not
correctly detected.
Based on the assessment of
benefit and harm, decide
whether you should activate
the HF device.
Pos : 37 /679-BOWA/HF- Ger äte/ARC 300/350/6 Bedi enung/UE_Bedienung @ 4\mod_1261040616100_6.doc @ 35033 @ 1
Содержание ARC 250
Страница 1: ...OPERATING MANUAL ELECTROSURGICAL UNIT...
Страница 2: ......
Страница 3: ...900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 3...
Страница 4: ...4 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...
Страница 5: ...900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 5...
Страница 8: ...Legend 8 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...
Страница 64: ...11 Technical specifications 64 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...
Страница 65: ...11 Technical specifications 900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 65...