31
9. PRODUCT STORAGE AND SHIPPING
Do not tilt or invert the console without first having
emptied the cooling liquid circuit.
Always empty the cooling liquid circuit before
longterm storage (2 weeks or more) or shipping to
avoid damage to the console.
Storage and transport conditions are specified in
the Technical Data section.
9.1. EMPTYING THE COOLING LIQUID CIRCUIT
1. Unplug all cables at the rear of the console.
2. Place the console on a flat, stable surface.
3. To remove the air vent plug, push the grey ring with
your left hand and simultaneously pull the plug.
Figure 76
4. Make sure that the metal locking device is in the
down position.
Figure 77
5. Put the draining tube in a receptacle that is more than
600 ml in volume.
Figure 78
6. Connect the draining tube (supplied with the product)
to the outlet.
Figure 79
7. Tilt the console until the connector is in contact with
the flat, stable surface to fully empty the cooling liquid
circuit.