2. Remove the protective cap from the console.
Figure 32
3. Connect the handpiece to the console.
Figure 33
4. Pay attention to the orientation of the handpiece
connector.
The red dot must be on top for proper alignment.
5. Make sure that the handpiece cord does not touch
the floor and is not compressed or squeezed in any
way that might impede circulation of the cooling liquid.
6. The handpiece connection to the console is maintained
by a mechanical lock. During use, the lock icon (orange
handpiece activation icon) remains illuminated.
Do not exceed the maximum number of usage
cycles for the handpiece as specified in the
Technical Data section.
3.9. INSTALLING A PROBE ON THE HANDPIECE
1. Select the appropriate probe.
Refer to the Probe Compatibility Table section.
2. Use the wrench to firmly tighten the appropriate probe
on the handpiece.
Multiuse Torque wrench
Figure 34
3.10. CONNECTING THE POWER CORD
Connect only to a FI protected mains power supply
(FI = Residual current protection).
To prevent damage to the console, make sure that
its rated voltage meets the local line voltage.
Connect the power cord to the power socket at the rear
of the console.
Figure 35