19
5.2. PROBE INSERTION
Do not start treatment without ensuring that a
back-up probe is available.
To avoid bending the probe, make sure that the
probe and the endoscope are aligned.
1. Introduce and position the probe inside the endoscope.
2. The probe shall be in contact with the stone.
3. Make sure that the operation is performed with
continuous endoscopic vision.
5.3. TREATMENT SETTINGS
1. The probe is automatically recognized by the
handpiece to configure the console parameters for
each probe type.
2. If a new probe is connected, the system automati-
cally set the recommended treatment parameters of
this probe.
Figure 50
3. All probe and handpiece usage information are
automatically recorded in the console (number of
uses, time of use, etc.).
4. According to the type of treatment, two pre-settings
are available:
• Hard Stones Treatment,
• Soft Stones Treatment.
5. You can also set each parameter manually.
Refer to the following sections:
- Custom Settings,
- Hard Stones Treatment Settings,
- Soft Stones Treatment Settings.
5.3.1. Custom Settings
1. From the
STAND BY
screen, press the
START
button.
Figure 51