13
3.7. INSTALLING THE SINGLE-USE FLUID
MANAGEMENT SYSTEM SET (OPTIONAL) AND
REPLACEMENT POUCH
1. Suspend the two fluid pouches, on the cart or on an
IV pole, at a level that is lower than the console.
Figure 27
2. Connect the fluid management system input tube (A)
to the stone catcher output tube connector.
Figure 28
3. Close clamp (B) of one pouch to fill the first pouch.
Clamp (C) stays open.
A
B
C
Figure 29
4. When the open pouch is filled, open the closed clamp
(B) first.
5. Close the open clamp (C) (adjacent to the filled pouch).
6. The filled pouch can be exchanged for a new empty
pouch, using the Luer-lock connection.
Figure 30
3.8. CONNECTING THE STERILIZED HANDPIECE
TO THE CONSOLE
Make sure that the handpiece connector is dry be-
fore connecting it to the console.
1. To remove the protective cap from the handpiece cord,
hold the metal part of the handpiece cable connector
and push up on the cap using your thumb and index
finger.
Figure 31