Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The SWISS LITHOCLAST® TRILOGY
is intended for use in the electromagnetic environment specified below. The
customer or the user of the SWISS LITHOCLAST® TRILOGY
should ensure that it is used in such an environment.
38
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 2, ± 4, ± 8, ± 15 kV air
± 2, ± 4, ± 6, ± 8 kV contact
± 2, ± 4, ± 8, ± 15 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic
material,
the
relative humidity should be
at least 30%.
Electrical fast
transient / burst
IEC
61000-4-4
± 2 kV, 100 KHz for power
supply lines
±1 kV, 100 KHz for
input/output lines
± 2 kV, 100 KHz for power
supply lines
±1 kV, 100 KHz for
input/output lines
Mains power quality should
be
that
of
a
typical
commercial
or
hospital
environment.
Surge
IEC 61000-4-5
± 0.5, ± 1 kV line(s) to line(s)
±0.5, ±1, ± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality should
be
that
of
a
typical
commercial
or
hospital
environment.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location
in
a
typical
commercial
or
hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 % UT (>95 % dip in UT) for
0,5 cycle
40 % UT (60 % dip in UT) for 5
cycles
70 % UT (30 % dip in UT) for
25 cycles
0 % UT for 0,5 cycle at 0°, 45°,
90°, 135°, 180°, 225°, 270°
and 315°
0 % UT for 1 cycle single
phase
<5 % UT (>95 % dip in UT) for
5 s
0% UT for 250 cycles
<5 % UT (>95 % dip in UT) for
0,5 cycle
40 % UT (60 % dip in UT) for 5
cycles
70 % UT (30 % dip in UT) for
25 cycles
0 % UT for 0,5 cycle at 0°, 45°,
90°, 135°, 180°, 225°, 270°
and 315°
0 % UT for 1 cycle single
phase
<5 % UT (>95 % dip in UT) for
5 s
0% UT for 250 cycles
Mains power quality should
be
that
of
a
typical
commercial
or
hospital
environment. If the user of
the SWISS LITHOCLAST®
TRILOGY
requires
continued operation during
power mains interruptions, it
is recommended that the
SWISS
LITHOCLAST®
TRILOGY be powered from
an
uninterruptible
power
supply or a battery.
UT is the a.c. mains voltage
(100-240) prior to application
of the test level.
Table 14