5-1-2 LOAD CELL
Connection
(1) (2) (3) (4) (5)
Terminal No.
Name
Contents
1
SIG -
Load Cell Input Voltage (-), Blue wire
2
SIG +
Load Cell Input Voltage (+), Green wire
3
EXC -
Load Cell Voltage (-), White wire
4
EXC +
Load Cell Voltage (+), Red wire
5
SHIELD
Load Cell Grounding
Warning
Load cell output is very weak thus load cell cable shall be
separated and then installed.
Please use shield cable for load cell cable.
In case of extending the load cell cable, use of load cell cable
of our company is recommended.
Shield cable of load cell cable shall be connected to BS‐235
terminal only.
In case of sealing, road cell input cable SIG+ (green) and SIG‐ (blue) cable shall be connected to No.1 and
No.2 respectively
.
There may be abnormality in relay operation and option output when the weight is displayed in (‐) value.
Please check whether or not color of cable for each manufacturer’s and load cell model is
different or not before the connection.