12-1-3 Optional output
(DAC, RS-232C, RS-422/485, BCD) selection mode
It is a mode to select optional output.
It shall be selected in case of equipment with change in option.
In case of using analog output voltage (current) and serial output (RS-232C, RS-422/485),
it shall be changed into DAC and RS-232 or RS-485 respectively.
Select optional output with the
use of No.2 and No.3 key.
Select either DAC or RS-232 or
RS-485 output.
Selection shall be made based
on equipped option.
It moves onto next stage when
pushing No.4 key after selection
and push No.1 key for
cancellation.
12-1-4 Analog output
(DAC) voltage (current) check
It is a mode to check analog output voltage (current).
Analog high output setting value is displayed to meter display unit.
Check output voltage (current) with the equipment of digital multi meter
to the outside.
When there is change in analog high output setting value, output of voltage (current) is conducted
based on display value.
It is a mode to check analog
output voltage (current).
Digital multi meter shall be
quipped to output terminal.
Output of voltage (current) is
conducted from analog output
terminal based on its value when
the change in value is conducted
with the use of No.2 and No.3 key.
It moves onto next stage when
pushing No.4 key and push No.1
key for the cancellation.
Use the check mode after selecting DAC in
12-1-3 Optional output.
For range of setting value change, setting is available within ‐199999 ~ 999999.
(Initial value is linked to analog Hi output setting value. Initial value is set as 10.000 at the forwarding)
In case of checking
DAI analog output (current), output is no conducted even when (-) value is set.
Содержание BS-235