After ‘
LCCAL
’ display,
Input load cell rated output
using No. 2 and No.3 key and
Push No.4 key.
ex) 2.0000 mV/V
After ‘
CAPA
’ display,
Input rated capacity of load cell
with the use of No.2 and No.3
key.
ex) 20.000 kg
Digital input is completed when
pushing No.4 key and it returns
to measurement mode.
Step 3. Zero Calibration
Push No.2 key while pushing
No.4 key.
Zero calibration is
completed.
If zero calibration is performed in an unstable environment, measurement errors may occur.
Warning
The above method is described as an example, so you need to set it up after checking the load cell
installed in the field.
Although the setting range of resolution is 1/30000 or lower but display is conducted even when it
exceeds 30000.
Error may generate for digital calibration mode.
Conduct regular check on measurement and conduct calibration upon necessity.
Calibration under instable environment may cause the measurement error.