40
■
r (range) ON
relay motion mode :
r1.
r
on, r2.
r
on, r3.
r
on
: Indicates the result of judgment and turns on.
If Lo judgment value
≦
the measured value
≦
Hi judgment value.
L3 Hi value
L3 Lo value
L2 Lo value
L2 Hi value
L1 Hi value
L1 Lo value
[ HI ]
[ OK ]
[ LO ]
LO (L1) Relay Output
OK (L2) Relay Output
HI (L3) Relay Output
Weight
Time
0
L1 OFF
L2 OFF
L3 OFF
L3 OFF
L2 OFF
L1 OFF
OFF
OFF
OFF
L1 ON
L2 ON
L3 ON
L3 ON
L2 ON
L1 ON
OFF
OFF
OFF
OFF
OFF
OFF
Example )
Range ON
mode
rLY.1H
L1 Hi side judgment value
1000
1000 within the limits will activate a relay
OFF
900 within the limits will activate a relay
ON
900 beyond the limits will activate a relay
OFF
rLY.1L
L1 Low side judgment value
900
rLY.2H
L2 Hi side judgment value
2000
2000 within the limits will activate a relay
OFF
1900 within the limits will activate a relay
ON
1900 beyond the limits will activate a relay
OFF
rLY.2L
L2 Low side judgment value
1900
rLY.3H
L3 Hi side judgment value
3000
3000 within the limits will activate a relay
OFF
2900 within the limits will activate a relay
ON
2900 beyond the limits will activate a relay
OFF
rLY.3L
L3 Low side judgment value
2900