6. CALIBRATION
There are three types of calibration mode.
6-1 Calibration Mode
6-1-1 Mode
■ Digital calibration mode
Calibration shall be conducted not by using the actual load but entering rated output and
rated capacity of sensor with key.
■ Actual load calibration mode
It is calibration mode to adjust zero and span with the use of actual load.
■ Linearization calibration mode
It is calibration mode to conduct non-linear calibration that reduces the measurement error
through calibration of maximum 4 points excluding zero with the use of actual load.
6-1-2 How to select calibration mode
1.
It turns into the mode selection status when key is pushed while pushing key is
pushed at measurement status.
2.
Mode change and setting value change is available when key or key is pushed at
mode selection status.
3.
Push key in order to move onto next stage after saving the setting value at mode
selection status.
4.
Push key to cancel the setting and return to measurement mode for mode
cancellation.
When you push No.3 key
while pushing No.2 key,
entry to calibration mode is
available.
Change the mode with
No.2 or No.3 key.
(
d-CAL→A-CAL→L-CAL
)
After selecting the mode,
Press No.4 key to proceed
with calibration.
Push No.1 key for mode
cancellation.