34
8-3-3 Method by external hold input terminal
Hold is commenced when external hold input terminal is turned on (contact point input).
At the moment, external hold input terminal shall maintain on status.
Hold is cancelled when input is conducted under off status.
8. HOLD
6. COM1
ON
OFF
BS-235 Inside
SWITCH or RELAY
OPTION
SI
G
+
SI
G
-
SH
D
4
1
3
2
6
5
7
8
9
10 11 12 13 14 15
ZE
R
O
COM
1
INPUT
RELAY OUTPUT
(L
1)
LO
HOL
D
COM
2
HI
OK
(L
3)
(L
2)
POWER
F.G
GN
D
L (
+)
AC 85~240V
DC 10~30V
1
2
3
4
Hold section (flickering of hold LED)
External hold input
Operation is not conducted when hold mode is not set as edge or peak hold from function.
Switch shall be zero voltage input of machine contact point and semiconductor contact point.
Zero calibration is not conducted under Hold status.
ON, OFF time shall be secured for 10ms or longer.
Please use it within rated load.
8-3-4 Method by serial output (RS-232C/422/485) command
Hold is commenced by hold on command through communication and hold is cancelled
by hold off command.
Please refer to
11. Serial Output
for detailed contents.
1.
Hold ON command
Hold is commenced
.
Command Example
ID
Command
1
2
3
4
5
<STX>
0
1
H
<ETX>
※ BS-235 Format
2.
Hold OFF command
Hold is cancelled
.
Command example
ID
Command
1
2
3
4
5
<STX>
0
1
C
<ETX>
※ BS-235 Format