BT-770 Operation manual
107
P/N : 770-ENG-OPM-EUR-D03
Bistos Co., Ltd.
2018.05
HR
Measuring range
Adult: 15 ~ 300 bpm
Pediatric: 15 ~ 350 bpm
Resolution
1 bpm
Heart rate measurement
error
± 1 bpm or ± 1%, whichever is greater
Heart rate measuring
accuracy and response to
irregular rhythm
Ventricular bigeminy
80 ± 1 bpm
Slow alternating ventricular bigeminy
60 ± 1 bpm
Rapid alternating ventricular bigeminy
120 ± 1 bpm
Bidirectional systoles
90 ± 2 bpm
Time to alarm for
tachycardia
1 mV, 206 bpm Ventricular tachycardia
<10 s
0.5 mV, 206 bpm Ventricular tachycardia <10 s
2 mV, 206 bpm Ventricular tachycardia
<10 s
2 mV, 195 bpm Ventricular tachycardia
<5 s
1 mV, 195 bpm Ventricular tachycardia
<5 s
4 mV, 195 bpm Ventricular tachycardia
<5 s
HR Alarm
HR upper limit
Adult: 16 ~ 300, 1 bpm step
Pediatric: 16 ~ 350, 1 bpm step
HR lower limit
Adult: 15 ~ 299, 1 bpm step
Pediatric: 15 ~ 349, 1 bpm step
TEMP
Standard compliance
ISO 80601-2-56:2009
Measurement method
Thermistor
Operating mode
Direct mode
Measuring range
0
℃
~ 50.0
℃
(32
℉
~ 122.0
℉
)
Resolution
0.1
℃
Measurement accuracy
± 0.3
℃
Number of channel
2
TEMP Alarm
T1/T2 upper limit
0.1
℃
~ 50.0
℃
,
0.1
℃
/
℉
step
T1/T2 lower limit
0
℃
~ 49.9
℃
,
0.1
℃
/
℉
step
TD upper limit
0
℃
~ 50.0
℃
,
0.1
℃
/
℉
step
RESP
Measurement method
Thoracic electrical bio impedance method
Measuring range
Lead RA-LA, RA-LL,LA-RL,LL-RL
Wave gain
X0.5, x1, x2
Respiratory impedance
range
0.2 ~ 3 Ω
Base line impedance
500 ~ 2 000 Ω
Scan speed
6.25 mm/s, 12.5 mm/s, 25 mm/s
Measurement accuracy
± 2 rpm
Measurement range
0 ~ 120 rpm