BT-770 Operation manual
12
P/N : 770-ENG-OPM-EUR-D03
Bistos Co., Ltd.
2018.05
instruction for use, in risk of external high voltage.
•
Do not connect any equipment or accessories that are not approved by the
manufacturer or according to IEC 60601-1 to the monitor. The operation or use
of non-approved equipment or accessories with the monitor is not tested or
supported, and monitor operation and safety are not guaranteed in such a
case.
•
Any non-medical equipment (such as the external printer) is not allowed to be
used within the patient vicinity (1.5m/6ft.).
•
Parts and accessories used must meet the requirements of the applicable
safety standards, and/or the system configuration must meet the
requirements of the medical electrical systems standard.
•
Ensure that the conductive parts of electrodes and associated connectors,
including neutral electrodes, do not come in contact with earth or any other
conducting objects.
•
Protection of ME EQUIPMENT against effects of discharge of a cardiac
defibrillator depends on use of proper cables.