BT-770 Operation manual
30
P/N : 770-ENG-OPM-EUR-D03
Bistos Co., Ltd.
2018.05
2.1.2 Placement requirements
Equipment installation must meet:
-
The left and right side of the monitor should have space more than 100 cm from the wall
-
Back on the monitor should have space more than 50 cm.
-
Ensure that the operating floor and the monitor have enough space for connecting the
accessory wires.
2.1.3 Power requirements
-
DC power supply adapter
Input: A.C. 100 V - 240 V, 50/60 Hz
Output: D.C. 15 V, 2.5 A
-
Built-in rechargeable lithium-ion battery: D.C. 11.1 V, 4400 mAh
2.1.4 Environmental requirements
The storage, transport and use of the monitor must meet the following environmental
requirements.
Operating
environment
Ambient temperature
5
℃
~ 40
℃
Relative humidity
30 % ~ 85 % (Non-condensing)
Atmospheric pressure
700 ~ 1060 mbar (hPa)
Transportation Prevent severe shock, vibration, rain and snow splashing during transport.
Storage
The packaged monitor should be stored in well-ventilated room with
ambient temperature -20
℃
~ 60
℃
, relative humidity 0 ~ 95 % (Non-
condensing), atmospheric pressure 700 ~ 1060 mbar(hPa), and without
corrosive gases.
The operating environment of the monitor should avoid noise, vibration, dust, corrosive or
flammable and explosive materials. In order to allow air flowing smoothly and achieve good
heat dissipation, at least 2 inches (5cm) clearance should be kept around the device.
When the device is moved from one environment to another, the device may have
condensation due to the differences in temperature or humidity. In this case, wait until the
condensation disappears before using the device.
WARNING
Ensure that the monitor is used under specified environment. Fail to do this, the
technical specifications declared in this manual may not be met and it may result in
damage to equipment and other unforeseen consequences.