16.2 Viewing software version information
You can view the software version through the following steps:
Select
【
Settings
】
Smart Hotkey→ “Settings” Menu;
Select “Monitor Info>>”→ “Monitor Info” menu;
“Monitor Info” menu displays the software version information of the monitor.
16.3 Maintenance plan
The following tasks can only be done by qualified service personnel of Bistos. When the
following maintenance is needed, please contact your service representative. Before testing or
maintenance, clean and disinfect the device.
Inspection / Maintenance Item
Frequency
Check the safety according to IEC
60601-1
At least once every two years, after replacing the
power supply or the monitor falls down.
Check all monitoring or measuring
functions not listed
At least once every two years, or when you suspect
that the measured value is not accurate.
NIBP leakage test
At least once every two years, or follow hospital
regulations
NIBP calibration
At least once every two years, or follow hospital
regulations