BT-770 Operation manual
9
P/N : 770-ENG-OPM-EUR-D03
Bistos Co., Ltd.
2018.05
0.1 General precautions, warnings and cautions
•
Examine the patient monitor and any accessories periodically to ensure that the
cables, adapter cords and instruments do not have visible evidence of damage that
may affect patient safety or performance. The recommended inspection interval is
once per week or less. Do not use the patient monitor if there is any visible sign of
damage.
•
Only the DC power adapter supplied with the BT-770 is approved for use with the
device.
•
Do not attempt to service the BT-770 patient monitor. Only qualified service
personnel by Bistos Co. Ltd. should attempt any needed internal servicing.
•
Perform periodic safety testing to insure proper patient safety. This should include
leakage current measurement and insulation testing. The recommended testing
interval is once per year.
•
If the hospital or healthcare institutions using this device fail to implement a
satisfactory maintenance schedule, it will result in device failure and may endanger
the patient’s safety.
•
Use the patient monitor under the conditions specified in this operation manual.
Beyond the conditions, the patient monitor may not function properly and the
measurement results may not accurate and may result in device failure or
endangering the patient’s safety.
•
Do not operate the BT-770 patient monitor if it fails to pass the power on self-test
procedure.
•
During the operation, do not disconnect any cable.
•
The BT-770 patient monitor is intended to be used by clinical professionals or trained
doctors, nurses or laboratory assistant.
•
Do not service and maintain or clean the device including accessories while in use
with a patient.
•
Using the device to one patient at a time.
WARNING
•
Thoroughly read and understand the manual prior to use of the BT-770. Failure
to do so could result in personal injury or equipment damage.
•
The device is intended for clinical patient monitoring, and only trained and
qualified doctors and nurses should use the device.
•
The alarm volume, upper and lower alarm limits should be set according to the
actual situation of the using environment. Do not just rely on audio alarm