BT-770 Operation manual
88
P/N : 770-ENG-OPM-EUR-D03
Bistos Co., Ltd.
2018.05
11.8 Clean and disinfection method of NIBP cuff
If necessary, NIBP cuff and NIBP extension tube can be cleaned and disinfected together
without separated
11.8.1 Cleaning method
1.
Prepare enzyme cleaning agent, distilled water and 10% solvent, respectively in different
spray bottle.
2.
Sprinkle cleaning agent on NIBP cuff, connector and extension tube, keep 1 minute for the
dry stains.
3.
Use a soft cloth to wipe smooth face. Use soft hair brush to brush visible stain and
irregular surface
4.
Rinsed with copious amounts of distilled water.
NOTE
Please be especially careful to clean the air ball and control valve of whole air
system. Do not allow any liquid entering into reversing valve and saturated valve.
Don’t use a soft cotton ball and fiber to clean this accessory because they will stick
on the cuff and extension tube.
11.8.2 Disinfection method
1.
Sprinkle bleach solution (Formula: the proportion of water and bleaching powder to 1:10)
then keep 5 minutes
2.
Wipe off excess bleach solution and elute with distilled water again
3.
Natural dry cuff
11.9 Alarm setup
In “NIBP Setup” menu, select “Alarm Setup >>” to enter “Alarm Setup” interface, and set NIBP
alarm switch, alarm level, upper and lower alarm limit. See 6.5 Alarm Setup for detailed setting
method.