BT-770 Operation manual
23
P/N : 770-ENG-OPM-EUR-D03
Bistos Co., Ltd.
2018.05
1.5 Description of monitor
Figure1-4: Front view
Name
Description
1
Alarm indicator
Indicates the priority of physiological alarm and technical
alarms in different colors and flashing frequencies.
-
High priority: Red, fast flashing (1.4 - 2.8 Hz)
-
Medium priority: Yellow, slow flashing (0.4 - 0.8 Hz)
-
Low priority: Yellow, constant on
2
Display area
Display the waveform and measured value
3
[Power]
-
Power On: Press down the key more than 2 seconds.
-
Power Off: Press down the keys more than 2 seconds
and the system will display the alarm message “The
system will shut down 3 seconds”.
4
Battery indicator
-
On: The battery is being charged or has been fully
charged.
-
Off: The battery has not been installed.
-
Flashing: The monitor is being powered by the
battery.
5
DC power indicator
Turned on when the monitor is being powered by the
adapter.
6
[Alarm reset]
To reset the alarm condition.